A Service A Plus LLC. ASAP Biomedical Engineer Support

Your #1 Second Source for everything for all your Beckman Coulter Hematology Units

Tech Support 573-642-2593 Web site www.coulterserv.com

 

Fax 573-642-0896 e-mail info@coulterserv.com

 

MAXM /HMX / LH 500 Special Tubing and Routing Chart

 

1018010

Tubing, special, sample line, 14.6 in. long

Routing: At CBC module, from BT2-7 to SF1

1018301

StabiLyse line (delivery to mixing chamber) Without Black I Beam Tubing

6857275

StabiLyse line (delivery to mixing chamber) With Black I Beam Tubing

Routing: At Mixing module, from MC1-5 through PV53 (normally closed) to FF173-1

 

6858077

Tubing, special, 7.0 (7.6) in. long

Routing: At Pump module, from FMT1-3 to CV79-2

At Main Diluter module -

From MF9-3 to FF205-2 (6 Vacuum Restrictor Line)

6859002

Pre-prep/LATRON control delivery line, 6-in. long

Routing: At BSV module, from FT31-2 to FF29-2.

From Green stripe of FF 42 to top of PV11-2

6859003

Pre-prep/LATRON control delivery line, 18-in. long

Routing: From BSV module FF37-2 to Mixing module MC1-3

From bottom of PV11-2 to top of mixing chamber

6859115

Tubing, special, 4.5 in. long, sheath restrictor #1 (For adjusting Diff Timing)

Routing: At Mixing module, from FF176-1 to FF177-1

6859118

Tubing, line choke, 4.75 in. long. (CBC lytic reagent restrictor)

Routing: At CBC module, from FU7-2 to FY8-2

Adjusts CBC Lyse delivery timing.

6859267

Tubing, line choke, 7.5 in. long (CBC lytic reagent restrictor)

Routing: At CBC module, from FU7-2 to FY8-2

Adjusts CBC Lyse delivery timing.

7000095

Rinse line, 4.5 in. long

Routing: At BSV module, from BSV WM6 to Peltier module, PLB1-2

7000096

Rinse line, 4.0 in. long

Routing: From BSV module Peltier module, PLB1-3, to Mixing module FF172-2

7000097

Sample line, 11.0 in long (Sample Delivery to Mixing Chamber)

Routing: From BSV module, BSV WM11, through Mixing module PV51 (normally open) to MC1-4

7000100

Tubing, special, 6.0 in. long (6 Sample Delivery Line)

Routing: From Mixing module MC1-7 through PV55 to Flow Cell module FC1-1

7000101

Tubing, special, 3.5 in. long sheath restrictor #1 (For adjusting Diff Timing)

Routing: At Mixing module, from FF176-1 to FF177-1